Signal Viewer

The Signal Viewer panel is a real-time multi-channel oscilloscope. It displays scrolling waveforms from the acquisition stream in µV, using GPU-accelerated rendering for smooth display even at high channel counts.

View Modes​

ModeDescription
OVERLAYAll channels superimposed on a single axis. Good for comparing amplitude across channels.
MATRIXEach channel in its own grid cell. Good for monitoring individual channels.
MEA 8×8Channels at physical electrode positions. Requires MEA Mapping.

Controls​

ControlDescription
View modeOVERLAY / MATRIX / MEA 8×8 toggle
LayoutAuto / 4×N / 6×N / 8×N / MEA 8×8
Display modeScroll: the window moves with the signal. Sweep: a cursor writes over the previous pass, like an oscilloscope.
DecimationClassic: keeps one sample out of N (a narrow spike can be missed). Min/Max: keeps the minimum and the maximum of each bucket, so spikes are always visible.
WindowDisplay window in samples: 1,000–50,000 in Scroll mode, up to 100,000 in Sweep mode (steps of 1,000)
RefreshUpdate interval: 20–500 ms (steps of 10)
Y min / Y maxAmplitude range in µV. Click AUTO Y to auto-scale.
Channel selectorPer-channel enable checkbox and color picker
PerformanceGreen = on budget / Yellow = near limit / Red = overloaded

Zooming on Several Channels​

In MATRIX view, click a cell to open a zoom window on that channel (see below). To compare several channels in the same window, click ADD in the zoom window header: the matrix returns with the message Select the signals to zoom. Click cells to tick or untick channels (a counter shows how many are selected), then press OK; Cancel or Esc goes back to the zoom window with the previous selection. Leaving the MATRIX view closes the zoom.

Per-Channel Modal​

Click any cell in MATRIX view to open the per-channel modal:

  • Time view — waveform oscilloscope
  • FFT view — frequency spectrum (1k, 2k, 4k, 8k, 16k sample FFT)
  • Dual frequency band slider for highlighting frequency ranges

Tips​

  • Start with MATRIX view to identify active channels, then switch to MEA 8×8 for spatial context.
  • If the performance indicator turns red, reduce the window size or disable unused channels.
  • Use AUTO Y after changing channels or electrode configurations to re-scale the display.