Skip to main content

Raster

The Raster panel displays a spike raster plot. Each row is one acquisition channel (0 to N−1, where N is the device channel count). The horizontal axis is time. Each tick mark is a detected spike. Provides a fast overview of multi-channel spiking activity.

Controls​

ControlDescription
WindowTime axis duration: 1–10 seconds
RefreshUpdate interval: 100–1000 ms
Band-passEnable/disable signal band-pass before detection
f_low HzHigh-pass cutoff (default: 300 Hz)
f_high HzLow-pass cutoff (default: 3000 Hz)
k×RMSThreshold multiplier on RMS (default: 5.0)
|y| thresholdDetect by absolute amplitude instead of k×RMS
BiphasicBiphasic spike detection
FusionOR: accept either biphasic component / AND: require both
d_min msRefractory period (default: 0.75 ms)
RMS win msWindow for σ estimation (default: 1000 ms)

Interactions​

  • Hover on the canvas — tooltip shows channel number and time of the nearest spike
  • Click on a tick — opens a waveform modal with a ±1 ms snippet around that event
note

The Raster panel uses its own independent software-side spike detection, separate from the firmware-side SPIKE ON command / Studio block. Both can run simultaneously.